I. Overview. 
The JNYQ-O- 15 laser analyzer adopts the principle of tunable laser absorption spectroscopy ((Tunable Diode Laser Absorption Spectroscopy, referred to as TDLAS), which can measure the concentration of specific gases in the process gas composition, including CO, CO2, O2, H2S, HCL, HF and so on. 
The system has the characteristics of high sensitivity, fast response, no interference from background gas, non-contact optical measurement and so on. 

2.Principle
 
The laser process gas analysis system is based on the international leading semiconductor laser absorption spectroscopy technology (DLAS), namely "single line spectrum" measurement technology. In the system, a modulable semiconductor laser is used as the light source, and the laser frequency is modulated by modulating the working current intensity of the semiconductor laser, so that the laser scanning range is slightly larger than the single absorption line of the measured gas. As a result, the laser beam of a specific wavelength emitted by the semiconductor laser is absorbed by the measured gas when it passes through the measuring tube, resulting in the attenuation of the laser intensity. Therefore, the system makes use of the different characteristic absorption lines of different gas components and the Beer-Lambert relationship between gas concentration and infrared or laser absorption spectra, and the concentration of the measured gas can be obtained by detecting the absorption size of the absorption lines (that is, laser intensity attenuation information). 

3. Characteristics. 
Free from cross interference of background gases. 
It can be used in bad working conditions such as electric coke capture. 
Low maintenance cost. 
Free from dust and window pollution. 
It is not disturbed by the change of environmental parameters of the gas being measured. 
Integrated design, compact structure and high reliability. 
Modular design, can replace all functional modules in the field, including laser module. 
High degree of intelligence, convenient operation and maintenance. 
There is no sampling preprocessing system, and the structure of the system is simple. 
The response speed is fast, the response time is not more than 1s, and the process control is guaranteed in time. 
No sample gas emission, environmental protection and no pollution. 
The process gas is analyzed directly, and the measurement accuracy is high.
  
4. Technical parameters
  
Gas  Lower monitoring limit Measuring range
O2 0.01%Vol. 0~1%Vol.,0~25%Vol.
NH3 0.5ppm 0~10ppm0~100%Vol.
CO2 40ppm 0~1000ppm0~100%Vol.
CO 40ppm 0~1000ppm0~100%Vol.
HCl 0.1ppm 0~100ppm0~50%Vol.
HF 0.05ppm 0~20ppm0~100%Vol.
H2S 5ppm 0~100ppm0~100%Vol.
H2O 0.1ppm 0~100ppm0~100%Vol.
CH4 10ppm 0~200ppm0~100%Vol.
C2H2 0.5ppm 0~50ppm0~100%Vol.
C2H4 1.0ppm 0~100ppm0~100%Vol.