System structure
Online monitoring of volatile organic compounds (VOCs)
Volatile organic compounds (VOCs) are important precursors for the formation of O3 and PM2.5. Strengthening the governance of VOCs is an effective way to control O3 and PM2.5 pollution, as well as a powerful means to help enterprises save resources, improve efficiency, and reduce safety hazards. Since the implementation of the Action Plan for Air Pollution Prevention and Control, China has successively issued a series of documents, including the "Thirteenth Five Year Plan for the Prevention and Control of Volatile Organic Compound Pollution", the "Comprehensive Treatment Plan for Volatile Organic Compounds in Key Industries", the "2019 Monitoring Plan for Volatile Organic Compounds in the Air of Cities at or above the Prefecture Level", the "2020 Volatile Organic Compound Control Plan", and the "Notice on Organizing the Strengthening Supervision and Assistance of Summer Ozone Pollution Prevention and Control". China has also issued more than 10 emission standards for VOCs, including unorganized emission control, petroleum refining, petrochemical, pharmaceutical industries, and product quality standards that limit the content of VOCs in coatings, inks, adhesives, cleaning agents, and other industries, comprehensively strengthening the top-level design of VOC control. The TR-9300V fixed pollution source VOCs emission continuous monitoring system is a monitoring technology that uses full process high-temperature and wet sampling. It is equipped with advanced gas chromatography separation technology and FID detection method (GC-FID) to measure the components of volatile organic gases such as total hydrocarbons (THC), non methane total hydrocarbons (NMHC), benzene series, aromatics, esters, etc. in the exhaust gas, and transmit the measurement data to the environmental protection department. The internal sampling, injection, and detection systems of the instrument all adopt high-temperature heat tracing method, which is conducive to accurate measurement of high concentration and high boiling point samples. For complex on-site application conditions, a structure that is easy to explosion-proof design is adopted, which can meet the monitoring needs of many industries such as petrochemicals, pharmaceuticals, automotive manufacturing, furniture manufacturing, semiconductor manufacturing, printing, spraying, etc. According to different monitoring requirements, this instrument has different customized ranges to choose from, meeting the monitoring needs of different users for VOCs.
![]() |
![]() |
| System structure | System diagram |
System composition
(1) Sample collection and extraction section: heating sampling probe, sample heat transfer pipeline, sample pretreatment unit (filtration, stable flow), sampling pump, etc.
(2) Sample analysis and testing section: Gas chromatograph (GC-FID).
(3) Other auxiliary testing parts: whole system calibration unit, exhaust emission unit, backflow control unit, generator discharge unit.
(4) Smoke emission parameter monitoring subsystem: The smoke temperature is measured using thermocouples or thermal resistors; The smoke pressure is measured using a pressure sensor; The flow rate or velocity of flue gas shall be measured using Pitot tubes and ultrasonic methods; Dry wet oxygen method for flue gas moisture content.
(5) Data collection, processing, and transmission part: collection, storage, recording, calculation and processing of sampling and testing data and system status parameters, as well as data uploading.
System characteristics
• The pre-treatment method complies with the measurement standards for fixed pollution source exhaust gases in the United States, the European Union, and China, and has high reliability;
• The system adopts the total heat method, with high temperature throughout the entire process from sampling to analysis, without the need for water removal; Unique insulation box design, the temperature of the sample gas outlet pipe wall can reach 120 ℃ -150 ℃, effectively avoiding sample loss;
• Flow path balance atmospheric pressure technology ensures the accuracy of detection results under different sample pressure conditions;
• Pre pumping circuit design, with a cyclic response period of ≤ 3 minutes, ensuring real-time and effective detection results;
• Equipped with automatic blowing function, it can automatically remove dust on the surface of the probe filter element, extending the service life of the probe;
• Equipped with full flow and automatic calibration functions, it meets the latest 1013 standard requirements for sampling full flow calibration;
• The chromatograph supports full automatic calibration without the need for on duty, minimizing manual maintenance;
• Using air as the carrier gas, there is no need to configure nitrogen cylinders, reducing maintenance costs by about tens of thousands of yuan annually;
• Users can choose explosion-proof design, adopting positive pressure explosion-proof cabinet design, which can be installed in explosion-proof areas, safe and reliable;
• Equipped with chromatographic injection backflushing function, it can avoid interference caused by residual sample gas and ensure the accuracy of analytical data.
Main uses and scope of application
The TR-9300V fixed pollution source VOCs emission continuous monitoring system is suitable for online continuous monitoring of VOCs concentration in fixed pollution sources in industries such as furniture, printing, automotive, leather, coatings, and petroleum.
The measurement objects are volatile organic compounds (VOCs, substances that are liquid at room temperature, solvents), such as gasoline, kerosene, light oil and other fuel oils, diluents, benzene, toluene, xylene, ethylbenzene, dichloromethane, trichloroethylene, tetrachloroethylene, alcohols, ethers, esters, aldehydes and ketones, amines, amides, NMP, nitriles, organic acids, organic silicon solvents (siloxanes), CFCs (fluorocarbons), and almost all organic solvent vapors.